Various concentrations of NH3 gases, ranging from 5 to 100 ppm, w

Various concentrations of NH3 gases, ranging from 5 to 100 ppm, were purged into the chamber in order to probe the sensing performance of the optimal Py-rGO sensor. As shown in Figure  8a, the plots of normalized resistance change versus time for the sensing selleck products device based on assembled Py-rGO upon exposure to NH3 PFT�� nmr gases with different concentrations were illustrated. The results revealed that the sensing device exhibited an excellent and highly reversible response to different concentrations of NH3 gases. When the NH3 gases were introduced into

the chamber, the resistance of the sensing device increased significantly over a period of 12 min, and the increase of the concentration of NH3 gas can result in the increase of the resistance of the device, and all of the resistance variations can be distinctly observed when the devices expose to the NH3 gas with the concentration

ranging from 5 ppb to 100 ppm. When the concentration of NH3 gas is 100 ppm, ca. 22% of the resistance change can be observed. As the concentration of NH3 gas decreases, the resistance change of the device decreases accordingly, and ca. 4.2% of the resistance change can be also observed when the concentration of NH3 gas was as low as 5 ppb. This is fascinating since the Py-rGO-based sensing devices exhibit much better response to NH3 gas than many other rGO-based devices Carbohydrate VX-689 [47, 48]. Furthermore, the relationship of response variation of the Py-rGO sensor as a function of NH3 concentration has also been studied as shown in Figure  8b. The sensing signal changed linearly with the concentration of ammonia when the concentration is above 50 ppb. The linear relationship between the response of Py-rGO and the concentration of NH3 is in

accordance with the work we reported before [29]. When the concentration is below 50 ppb, the sensing signal dropped rapidly (as shown in Figure  8b), which might be due to the PPy molecules covered on the surface of rGO sheets, and blocked the gas molecules interact with the rGO sheets, leading to a worse response to the NH3 gas molecules. Figure 8 The response performance of sensing devices based on assembled Py-rGO sheets. (a) Plot of normalized resistance change versus time for the sensing device based on assembled Py-rGO upon exposure to NH3 gas with concentrations ranging from 5 ppb to 100 ppm: a, 5 ppb; b 50 ppb; c, 10 ppm; d, 50 ppm; and e, 100 ppm. (b) Relationship of response variation of the Py-rGO sensor as a function of NH3 concentration. Furthermore, the sensor response exhibits an excellent recovery characteristic (as shown in Figure  8a). As illuminated with IR lamp together with flushed with dry air over the periods ranging from 134 to 310 s, the resistance of the device decreased and essentially recovered to the initial values.

Comments are closed.