Typical of isolation procedures, the recovery increased from a lo

Typical of isolation procedures, the recovery increased from a low of 50% at the lowest MV counts up to 80% at the highest counts. Scatter signals from MV isolated by ultracentrifugation ( Fig. 1B) were better resolved than those obtained from samples analyzed by direct staining of PFP or unwashed MV ( Fig. 6), which showed substantial populations of microparticles negative for all stains ( Fig. 6, red Compound Library chemical structure dots). Counts of MV were the same when isolated from either PFP or PPP stored at either -40 °C or − 80 °C for more than a year. Up to three freeze

thaw cycles of PFP had no effect on MV counts, irrespective of initial counts (Fig. 7). Once isolated, counts of isolated MV were stable during storage at room temperature for 3–4 days. However, a single freeze and thaw of isolated MV at either − 20 °C, − 40 °C or − 80 °C lowered the count by 10–15%. The assumption that the nominal TruCOUNT™ bead count is valid was verified by a cross-check with erythrocyte counts and a validated Coulter counter (Fig. 8). As the erythrocyte count in each sample increased above the order of the (constant) TruCOUNT™ bead concentration, the red blood cells (RBC) event rate increased in linear proportion Venetoclax mw to the RBC count while the TruCOUNT™ bead event rate declined. Because the TruCOUNT™ calibration is in the denominator

(Materials and methods), the calculated erythrocyte count showed a systematic increase (solid line/symbols) above that obtained with the Coulter counter (dashed line). Extrapolation of the linear increase to the erythrocyte count of zero intersected the count axis within 5% of the Coulter counter value, and showed a systematic error of + 10% when the count rate was 1000 times that of the TruCOUNT™ rate. Because analysis with other bead calibrators has been published (Robert et al., 2008), we analyzed mixtures of BD TruCOUNT™ beads (4.2 μm) with Beckman-Coulter Flow-Check (10 μm) beads for counts obtained by scatter and by fluorescence. In all cases, scatter and fluorescence data were congruent. Two lots of the

Flow-Check beads yielded counts of 50% of nominal or less when the TruCOUNT™ count rates were of the order of 20–30/s. At lower bead dilutions (higher count rates), CHIR-99021 nmr the BD beads yielded proportional counts whereas the Flow-Check beads were disproportionately undercounted. We did not investigate this disparity further. Distinct populations of circulating MV have been observed in a variety of disease conditions, often related to inflammatory processes (Zwaal and Schroit, 1997, Berckmans et al., 2001, VanWijk et al., 2003, Morel et al., 2006, Jayachandran et al., 2008 and Jayachandran et al., 2009). However, the potential for MV as biomarkers has been limited by inadequate validation and standardization of sample preparation, reagents and instrument parameters (Jy et al., 2004 and Lynch and Ludlam, 2007).

Leave a Reply

Your email address will not be published. Required fields are marked *

*

You may use these HTML tags and attributes: <a href="" title=""> <abbr title=""> <acronym title=""> <b> <blockquote cite=""> <cite> <code> <del datetime=""> <em> <i> <q cite=""> <strike> <strong>